Handheld Mass Monitors are Indoor/Controlled Environment Monitors BUT.....How to 'spot check' ambient air?
Service Bulletin
Aerocet 531, Aerocet 531S, 831 Mass Monitors
Subject: Using a K - Factor 13 MAY 2014
The accuracy of the Aerocet 531, 531S, or 831 Mass
Monitors depend highly upon the reflectivity of the particulate. The instrument
is calibrated in the factory using PSL particles. In many cases, real world
particulate can have a completely different characteristic than the calibration
aerosol, simply due to the reflectivity properties of the particulate. Coal
dust is significantly different in reflectivity characteristic than airborne
dust near an aluminum smelter. In many cases, these mass monitors will need a
K-factor set into them via the software. This is especially true when the
instrument is used outdoors for spot checks.
A K-factor is simply a multiplier. For example, if
you have a device that is reading only half of what it should be reading, you
normally apply a K-factor of 2 into it. The instrument multiplies the readings
by 2, and gives a more accurate comparison.
To set up a K-factor, you need to compare the mass
monitor data with a known good reference. This can be any mass monitor accepted
as having good data by the reporting agency. For example, a BAM 1020 would be a
known good reference since it has USEPA approval and similar approvall in most
countries.
The reference must be an instrument that is set
side-by-side with the handheld mass monitor. Using data from the nearest AQ
station is not close enough. The data on a rooftop on the other side of the
city could be very different than what you are monitoring. The instruments
should be properly collocated.
Below is an example of a K-factor comparison.
Mass Monitor Data (ug) Reference data (ug)
3 8
6 9
4 13
6 10
5 8
7 13
3 9
6 10
5 (Average) 10
(average)
Of course, you would use more points. The above is
an example only.
On our test, the reference is reading twice what the
mass monitor is reading. Thus, we would enter a K-factor of 2 into the instrument
setup screen through the USB port.
To do this, connect a USB cable between the device
and your computer. Open Comet and choose your station. The example below is for
the 831.
Look for the SETUP tab at the top of the terminal
window. It will appear as the picture below.
Using the up and down arrows, change the K-factor for each channel as
needed. Keep in mind that a K-factor of 0 cannot be used. The K-factor is a
direct multiplier. Thus, you would be multiplying your measured value times 0.
Please note: The Aerocet 531, 531S, and 831
Mass Monitors are indoor monitors. However, they can be used outdoors to spot
check a site.
For more information regarding K factors or about setting up your Mass
Monitor, call or email our service department at service@metone.com or call
541-471-7111 ext. 300.
Comments
Post a Comment