Skip to main content

BLACK CARBON MONITORS NOW DEPLOYED WORLDWIDE - MADE IN GRANTS PASS, OREGON, USA!


Met One Instruments BC 1050 and BC 1054


CONTACT SALES@METONE.COM

The Met One Instruments, Inc. BC 1050 Series, Black Carbon Monitors, measures the absorption of 
particulate matter onto filter tape continuously. The BC 1050 monitor operates at 2-wavelengths and 
the BC 1054 operates at 10-wavelengths. The BC 1050 offers the user the ability to monitor 
absorption at near-IR (880 nm) and near-UV (375 nm) wavelengths






simultaneously with a standard minimum time resolution of 1-minute. The BC 1054, multi-spectrum carbon monitor, measures the absorption at 375, 430, 470, 525, 565, 590, 660, 700, 880 and 950 nm with a standard time resolution of 1-minute. Optional
1-second time resolution is available for both the BC 1050 and the BC 1054.

Both monitors offer the following advanced features:
•   Active flow control between 2 and 5 LPM. Concentrations may be reported under either actual or standard flow conditions. Optional 16.7 LPM operation is  offered.
•   Compact flash storage of more than 1-years worth of 1-minute data standard. Expanded memory storage is available.
•   Built in filter-loading correction algorithm using the Park technique with user-selectable 
input criteria  standard.
•   “Data-Masher” for off-line filter loading correction provided with each unit   sold.
•   Cloud service is available.
•   Advanced diagnostics flow, optics, tape drive  operation.
•   Met sensor inputs.
•   Ultra-high sensitivity, sub-ng/m3 sensitivity on the hourly timescale.

Applications:
* Air Quality Surveillance Global Warming Studies Particulate Emissions Studies
* Near-Roadside Monitoring Visibility Studies
* Source Apportionment

All BC 1050 and BC 1054 monitors come equipped with active flow control; concentrations may be 
displayed under actual or standard flow conditions.

All BC 1050 and BC 1054 monitors are factory calibrated in a smoke chamber using the same method that almost 10,000 Met One Instruments beta attenuation mass monitors have been calibrated. This insures long-term and long-range reproducibility and traceability. All equipment returned for service is recalibrated against the factory reference standard before being returned to the field thereby ensuring continuity of data collection.

Specifications

The BC 1050 and BC 1054 monitors are manufactured at an ISO-9001 Facility.

BC 1050 and BC 1054 monitors have input and output options to enable the user to add other 
measurements and/or output data to cloud services or to memory sticks. The BC 1050 and BC 1054 will directly connect to a variety of products, such as the BAM 1020, discrete particle size and counting modules, or a wide array of meteorological sensor inputs.

PARAMETER  SPECIFICATION
Measurement Principle:    Filter-Based Multiple Wavelength Optical Absorption
BC 1054:  10 Wavelengths:
375, 430, 470, 525, 565, 590, 660, 700, 880, 950 nm
BC 1050:  2 Wavelengths:
375, 880 nm
Measurement Range:  <1 ng/m3 to >100,000 ng/m3 (effective) Display Resolution:             
        0.1 ng/m3
Limit of Detection:    < 8 ng/m3 with 1 minute sample time
        < 1 ng/m3 with 1 hour sample time
Measurement Interval:  1-second or 1-minute
Flow Rate:   2 or 5 LPM, user selectable, mass-flow 
controlled under actual flow  conditions.

16.7 LPM option available through EPA-style sampling  inlets
Pump Type:   Internal AC vacuum pump, other configurations available
Filter Tape:     Reinforced glass fiber
Power Supply: 12V DC. Universal 100-240 VAC 50/60Hz input, 12 VDC 
8.5A output supply included. Power Consumption: 50 W
Operating Temperature:   0 to +40°C
Ambient Temperature:    -30 to +50°C
Ambient Humidity Range:   0 to 90% RH, non-condensing
Data Storage:  One USB Flash Drive Port
Cloud service available: Stores more than 1 years worth of 1-minute data Expanded          memory  available
Data Collection Interface: Single serial output through RS-232, USB, or through Ethernet port. 1200 to 115200 baud. Compatible
Software:  Comet™ software and USB/Ethernet driver CD  included.
User Interface:  Menu-driven interface with 4x20 character backlit LCD display 
and dynamic  keypad. 
Mounting Options:   Bench top or equipment rack mountable. Rack mount hardware is   standard.
Inlet :  PM1, PM2.5  sharp cut cyclones (SCC) available at either 2 or 5 LPM Sampling  Rates             Through Flexible Tubing. 5 LPM PM2.5 SCC standard.
           For 16.7 LPM option EPA style PM10 inlet with PM2.5 SCC available.
Unit Weight:     Approximately 40 lbs (18  kg)
Unit Dimensions:      Height: 10.5” (26.7 cm)    Width: 17” (43 cm)  Depth: 15.8” (40  cm).





 























































JUNE 2016




























































1600 Washington Blvd., Grants Pass, OR 97526
Tel: 541.471.7111  | Fax: 541.471.7116
E-mail: sales@metone.com  |  www.metone.com